Brand:NANBEI
Model:FM-Nanoview 1000 AFM
Application:
Scan head and sample stage are designed together, strong anti-vibration performance;Precision laser detection and probe alignment device make laser adjustment simple and easy;Adapt servomotor to drive the sample approaching tip manually or automatically, to realize precision scanning area positioning;High-accuracy and large range sample transfer device allow to scan any interesting area of sample;Optical observation system for tip check and sample positioning;Electronic system is designed as modular and easy for maintenance and further development;Adopt spring for vibration isolation, simple and good performance.
Software
1. Two kinds of sampling pixel for choose: 256×256, 512×512;
2. Execute scan area move and cut function, choose any interesting area of sample;
3. Scan sample in random angle at beginning;
4. Adjust the laser spot detection system in real time;
5. Choose and set different color of scanning image in palette.
6. Support linear average and offset calibration in real time for sample title;
7. Support scanner sensitivity calibration and electronic controller auto-calibration;
8. Support offline analysis and process of sample image.
Main technical parameters
Item |
Technical data |
Item |
Technical data |
Operation modes |
Contact mode, friction mode, extended modes of Tapping, phase, MFM, EFM. |
Scan angle |
Random |
Sample size |
Φ≤90mm,H≤20mm |
Sample movement |
0~20mm |
Max. scan range |
X/Y: 20 um, Z: 2 um |
Pulse width of approaching motor |
10±2ms |
Resolution |
X/Y: 0.2 nm, Z: 0.05nm |
Optical system |
Magnification: 4X, resolution: 2.5 um |
Scan rate |
0.6Hz~4.34Hz |
Data points |
256×256,512×512 |
Scanning control |
XY: 18-bit D/A, Z: 16-bit D/A |
Feedback type |
DSP digital feedback |
Data sampling |
One 14-bit A/D and double 16-bit A/D multiple-channel simultaneously |
Feedback sampling rate |
64.0KHz |
PC connection |
USB2.0 |
Windows |
Compatible with Windows98/2000/XP/7/8 |