NT1810D Series Double Beam UV/VIS SPECTROPHOTOMETER

Characteristics of Products
● High Accuracy: Using grating micrometer level precision screw drive to ensure the accuracy of wavelength <±0.3nm;Transmittance of accuracy reaches ±0.3%.
● Convenience for use: 5.7 inch graphic LCD display. Spectrum curve be clear at a glance and easy to operate. There is quantitative analysis、qualitative analysis、 kinetics test、 DNA/RNA、 multi- wavelength analysis in special purpose testing program.
● Service life durable: Imported Halogen lamp ensures the light source for up two years of life, receiver life for 20 years.
●Wide application fields: Optional Equipped with Peltier/sipper unit, with Micro cell holder, with (5°incident angle) Reflectance measurement attachment etc. expands the application areas of more.
 
Function of products
●Photometric measurement: Convenience to measure samples in the specified wavelength absorbance and transmittance. Multi-wavelength detector is maximum simultaneous determination of 10 wavelengths.
●Quantitative measurement: Establish automatically standard curve, first- order 



Model NU-T1810D NU-T1810DA NU-T1810DS
Optical System Double Beam, Grating 1200 line/mm
Wavelength Range 190~1100nm
Spectral Bandwidth 1.8nm 1nm 0.5/1/2/4/5nm
Wavelength Accuracy ±0.3nm
Wavelength Repeatability ±0.1nm
Photometric Accuracy ±0.3%T;±0.0015Abs(0~0.5Abs);±0.003(0.5-1.0Abs);
Photometric Repeatability ±0.1%T;±0.001Abs(0~0.5Abs);±0.002(0.5-1.0Abs);
Stray Light ≤0.03%T, @220nm/360nm
Noise Level ±0.0008Abs
Stability ±0.0008A/h@500nm
Baseline Flatness ±0.0008Abs
Baseline Dark Noise 0.05%T